Replaces the sensor in EQ-W1000-LD:
(Analyzer & Sensor Housing NOT included) |
Available in two versions: Glove Box and Furnace, please select from the Product Options above.
- Microfuel cell technology
- Accuracy: +/- 1% Full-Scale Range
- Repeatability: +/- 0.5% Full Scale
- Flow Sensitivity: 0.5 - 5 SCFH
- Pressure: 0.1 - 50 PSIG
- Temperature Coefficient: 2.54%/ ºC
- Operating Temperature: 0 to 50 ºC
- Storage Temperature: 0 to 45 ºC
- Recommended Gas Flow: 0.5 - 2 SCFH (Standard Cubic Feet per Hour) Furnace Sensor
- 0.5 - 5 SCFH Glovebox Sensor
- Humidity: 0 - 100% Relative; Non-Condensing
- Longevity: 17 - 25 Months
|
|
Sensor Types
|
Glove Box Sensor Provides the best response when used in enclosed environments where battery R&D is conducted; measures trace oxygen levels in purge gases (e.g. N2 and Argon). (Image for reference only, Glove Box system is NOT included)
|
Furnace Sensor Designed for CVD applications when acid and aggressive gases (e.g. CO2, HCI, HCN, and etc.) will be present. (Image for reference only, Furnace system is NOT included) |
| Shelf Life |
12 months
|
3 months |
| Weight |
1 lb |
| Notes & Warnings |
- The analyzer is designed for measuring trace oxygen in the presence of gas flow; it may not measure reliably when gas is static.
-
NEVERuse the oxygen sensor under vacuum conditions. The vacuum will cause irreversible damage to the sensor.
-
ONLY USE the furnace oxygen sensor with CVD furnace systems to prevent contamination and damage to the sensor.
- To prevent sensor damage, please use quick ON/OFF valves to isolate the sensor from open-air and maintain it with an inert gas at all times. Click the picture below to learn more about this setup:
- Due to the variations in electronics and sensor technology, minor deviations in O2 reading performance may be experienced when using O2 analyzers or sensors from a different brand, technology, and etc.
- The oxygen sensor is consumable and not covered by warranty, damage caused by misuse is also excluded from coverage.
|