Tube Dimensions
|
O.D. 85 mm I.D. 50mm Length: 1000mm |
| Inside Diameter: |
50 mm |
| Tube Length: |
1000 mm |
Max.Working Temperature without Load:
|
1100 °C |
Working Pressure & Temperature:
|
Maximum pressure inside the tube
|
Heating temperature
|
| Max. 1958 PSI (13.5Mpa) (140 Bar) |
800°C |
| Max. 1189 PSI (8.2 Mpa) (80 Bar) |
900°C |
| Max. 595 PSI (4.1 Mpa) (40 Bar) |
1000°C |
| Max. 362 PSI (2.5 Mpa) (25 Bar) |
1100°C (for <1hr) |
Warning:
- Please set up a pressure alarm point in the digital pressure display before the operation
- Must install a pressure regulator on the gas cylinder to increase pressure gradually.
- Service-life design for 2,000 hours at 800oC. The high-pressure vessel is a consumable part, please order replacement once found deformation
|
Pressure Gauge
|
- A digital Vacuum/Pressure Gauge is included with alarm function
- Pressure Limits for gauge: -100 Kp to 0 and to 30MPa (Vacuum to Atmosphere to Positive Gas Pressure)
|
| Installation: |
|
Water Cooling
|
- There are four water-cooling bracket blocks, two on each side.
- Two water cooling ports on each block.
- Illustration of linking all four half-water cooling blocks in series.
-
|
| Option Water Cilller |
- Must have a water chiller while heating this tube, or it will cause damage to the flanges.
- A water chiller is required, but not included, please click to order at extra cost.
-
|
Recommended Optional Tool
|
- Please use a torque wrench to make sure the equal forces to be applied to the bolts when installing the CF flange in MTI high-pressure furnace.
- Hex Bit Socket M14 of Torque wrench for CF flange installation CFTQ-M14
-
|