| Features |
|
| Length |
6" |
| Diameter of sheath |
0.32" ( 0.81 mm )diameter |
| Dimension |
|
| Positive Material |
Ni-Cr |
| Negative Material |
Ni-Al |
| Highest Temperature |
1200 oC |
| Lowest Temperature |
-200 oC |
| Application Note |
- If necessary, please check out K type thermocouple extension wire with mini male plug. Using compensating wires is equivalent to stretching them to keep them operating at a constant temperature. The advantages of using compensating cables are saving material costs and solving the problem of constant temperature, etc.
- There are 2 polarities for K type thermocouple: positive and negative. Please check polarity of K type thermocouple before use. Wrong polarity may cause temperature measuring error.
- It is also easy to test the polarity of K type thermocouples. The negative wire is MORE magnetic than the positive wire. Just put a magnet up to each wire. One will be more magnetic than the other.
|