| Purity |
Al2O3 > 99.8% |
| Density |
> 3.91 g /cm3
|
| I.D |
2-1/8"+/- 0.02 (53.5 +/- 0.5 mm) |
| O.D |
2.36"+/- 0.02 (60 +/- 0.5 mm) |
| Length |
790 mm (31.1") |
Detailed demension drawing |
|
| Max. working temperature |
1750 oC |
| Warning |
- Please don't use alumina tube under vacuum at temperature > 1500 oC
- Mcdanel strongly recommends keep ramp rate less than 100oC/ hour
- Please keep theral gradient less than 100oC/ inch to avoids cracking
- Parts should not be placed directly onto the inside surface of a hot furance tube incase of thermal shock
- Please handle the tube carefully. Warranty does not include tube (but only for machine MTI provided)
|
Application Notes
|
-
Please click here to learn how to use and maitain Alumina tube in MTI high temperature tube furnace.
- In order get longer tube service life, Please use quartz or alumia crucible to carry sample, instead loading sample directly inside tube, please click picture below to choose one crucible boat:
|
Inspect your delivery:
Please check package for damages before accepting delivery.
Should damages found, please keep all packaging materials, take pictures, and report immediately to carrier/MTI.