| Furnace casing |
|
| Total Power |
1.75 KW |
| Voltage |
AC 110V, Single Phase, 50/60 Hz
208V-240V version is available upon request
|
| Max. Temperature |
1200 °C (<1 hour) |
| Continuous Temperature |
1100 °C ( continuous) |
| Max. Heating Rate |
<= 10 °C/min |
| Processing Tube |
Quartz Tube: 50mm OD x 43mm ID x 600 mm Length |
| Heating Zone length |
8" (200mm) ( single zone ). The following picture shows the temp. distribution across the heating zone, please click to enlarge.
|
| Thermal block placement |
-
Two fibrous ceramic tube blocks are included for blocking heat radiation from tube chamber. (Ceramic blocks must be fully inserted into the furnace chamber before heating)
- Quartz blocks (optional) are ideal for cleaner and higher vacuum efficiency.
- You may order flange support to avoid tube rotating ( click Pic below right )
|
| Constant temperature zone Length |
4" (100mm) (+/-2°C) @ 900 °C
|
| Temperature controller |
PID automatic control via PWM solid relay with 30 steps programmable. |
| Temperature Accuracy |
+/- 1 °C |
| Vacuum Flange |
SS Vacuum Flanges (Left side comes with 1/4 barb fitting and TC feed through, right side comes with KF25 fitting vacuum gauge)
|
| Vacuum Level |
1E-2 (50 m-torr) by the included Double Step Rotary Vane VacuumPump (110V, 156 L/m) with Exhaust Filter & KF-D25 Inlet with Vacuum Bellows and KF25 Clamp
|
| Heating Element |
Fe-Cr-Al Alloy doped by Mo |
| Temperature Calibrator |
Thermocouples: K type 18" x 1/4"Dia with Alloy sheath and Male Plug for Calibration up to 1335°C Portable Digital Temperature Calibrator with USB and RS232 Interface & Software - NIST certificate
|
| Optional |
- This furnace can be updated to 60 mm quartz tube with quartz observation window as pic below at the extra cost
- For Vacuum to 10E-6 torr, you may consider turbopump with CF flange and one end open quartz tube.
( pls email for a quote) CF flange with one end closed tube can reaches 10-6 Torr by turbopump |
Laptop and software (Optional)
|
- Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30 days free trial) for immediate use.
-
Labview Based Temperature Control System (MTS01)enables users to edit temperature profiles, manage heat-treatment recipes, record and plot data for MTI furnaces.
|
Application Notes
|
- Tube furnaces with quartz tubes are designed for using under vacuum and low pressure < 0.2 bar
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
- Vacuum limit for all quartz tube furnaces: Vacuum pressures may only be safely used up to 1000°C
- The flow rate for gases should be limited to<200 SCCM (or 200ml/min) for reducing thermal shocks to the tube
- How to set up a ceramic tube and vacuum flange for MTI tube furnaces.
- For safety, the small tube furnace can be placed inside a fume hood. ( click the picture left to see )
|
| Certificate |
- NRTLs or CSA certification is available at extra cost.
|