Alumina Tube
|
-
O.D: 60 x I.D: 54 x L1000 mm (click underlined for specs)
- OD 80 mm and 100 mm tube is available in the product option
|
Furnace Structure
|
- The splittable design provides easy access to the tube replacement
- Double-layer steel casing with air cooling to ensure a low surface temperature < 60°C
-
High purity fibrous alumina insulation for Max. energy saving
-
The special refractory coating on the furnace liner and heating element to increase furnace life
-
Two sets of fibrous ceramic tube blocks are included for blocking heat radiation from inside the tube. (Ceramic blocks must be fully inserted into the furnace chamber before heating)
|
| Input Voltage |
208 - 240VAC SINGLE PHASE, 50/60Hz
|
| Power |
7 KW (40A Breaker)
|
| Heating Temperatures |
- Max. heating temperature: 1700°C, < 10 minutes (N2 or other inert gas must be filled inside the tube to resist deformation)
- Continuously working temperature: 1500°C (1600°C < 1hr)
- Heating rate: 10°C /min Max.
|
Heating Zone
|
- Total heating zone length: 400mm
- Constant temperature zone: 140mm (+/-1°C)
|
Heating Elements
|
- 16 pcs MoSi2 1800 grade Heating Element
|
Temperature Controller
Temperature controller interface
The optional PC control module
|
Standard temperature controller:
- Proportional–integral–derivative control (PID control) and auto-tune function
- 30 segments programmed with ramping, cooling and dwelling steps
- Built-in over-temperature alarm and thermocouple failure alarm
- +/- 1 ºC temperature control accuracy
- Default DB9 PC communication port (PC control module and software are sold separately.
Optional:
-
MTS-02 PC control module (also including cable and software) is an add-on option for PC control
- It is recommended to purchase the module together with the furnace. We install and test it before shipping
- It is recommended to purchase the laptop (with software pre-installed) from us to avoid software confliction
- Please note the module is not sold solely in the future
-
Eurotherm 3000 Temperature Controller is optional for an additional $950 including installation and testing. This package also includes the communication cable (RS485 - USB) and software CD. Click here to view the full spec
|
| Vacuum Sealing |
|
| Vacuum Levels |
- 10^-2 torr by the mechanical vacuum pump.
- 10^-5 torr by turbopump.
- Leak rate < 5 mtorr / min.
The vacuum pump is not included, please order separately at the related products section
|
| Power Cable & Plug |
- 10 feet length 10-3 AWG heavy duty UL approved power cable is included. Extra length cable can be ordered by click picture below-left
- The plug is not included. Pls, click compatible plugs to order or obtain one from a local hardware store.
|
| Compliance |
- CE Certified
- NRTL (UL61010) or CSA certification is available upon request at an extra cost
|
| Application Notes |
|
Warnings
Click here to learn more about the installation of a gas regulator
|
- NEVER open the furnace lid at temperature >100°C, doing so may thermal shock the processing tube and cause cracks.
- The tube furnaces with alumina tube are designed for use under vacuum and low pressure < 0.02 MPa
- Never use graphite crucible in the alumina tube.
-
Attention: A two-stage pressure regulator must be installed on gas cylinder to limit the-the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
- Never heat the furnace while the gas release valve is closed and a positive pressure condition exists within the furnace chamber. The pressure gauge must be used to closely monitor the chamber condition at all times during the heating process; please open the gas release valve immediately once the chamber pressure has reached over 3 PSI to prevent unforeseeable damages.
-
.The flow rate for gases should be limited to < 200 SCCM ( or 200ml/min )
- Vacuum pressure only is safely used up to 1500°C for alumina tube
|