Furnace Structure
|
- Double-layer steel structure with air cooling to keep the case temperature <65°C
- High purity fibrous alumina insulation with a Special coating to ensure max. energy saving
- Foam ceramic blocks with alumina coating are included, which is hung on top and bottom of flange to protect the O-ring from being damaged by heat radiation.
- The wheel is installed on the split top cover.
|
|
Power
|
4.8KW (30Amps circuit breaker is recommended) |
|
Voltage
|
AC 208-240V Single Phase, 50/60 Hz |
|
Max. Temperature
|
1200°C (<1 hr) each zone |
|
Continuous Temperature
|
1100°C for each zone |
|
Max. Heating Rate
|
<= 20°C/min. |
|
Temperature Accuracy
|
+/- 1°C |
|
Tube Size and Materials
|
- One 1200mm length quartz tube is included
- OD: 50 mm, ID. 44 mm
- Foam ceramic blocks with alumina coating are included, which is hung on top of the flange to protect O-ring damaged by heat radiation
- The furnace can be modified for Ø50 - Ø130 mm quartz tube.
|
|
Heating Zone Length
|
- Heating Zone: Total: 610 mm (152.4 mm +304.8 mm + 152.4 mm )
- Constant Temp. Zone: 480 mm within +/- 2 oC
|
|
Temperature Controller
|
- MET certified 30 programmable segments for precise control of heating rate, cooling rate, and dwell time.
- Built-in PID Auto-Tune function with overheating & broken thermocouple broken protection.
- Over-temperature protection and alarm allow for operation without an attendant.
- RS485 Communication Port.
- You may upgrade the controller to Eurotherm 3000 for +/- 0.1oC accuracy
|
Vacuum Sealing Flange
|
|
|
Vacuum Level
|
-
10E-2 Torr by a mechanical vacuum pump
-
10E-5 Torr by molecular pump
-
Leak rate < 5 mTorr / min.
|
|
Heating Element
|
1300°C Fe-Cr-Al Alloy doped by Mo
|
|
Laptop, software (Optional)
|
- Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30 days free trial)
-
Labview Based Temperature Control System (MTS01)enables users to edit temperature profiles, manage heat-treatment recipes, record, and plot data for MTI furnaces.
- A wireless remote control provides up to 300 meters of operating range.
|
| Compliance |
- CE Certified
- NRTL or CSA certification is available at extra cost.
|
|
Application Note
( We may help you to build customized flange to meet your application at extra cost )
|
-
- Please click the picture below to learn how to quench sample in gas flow and vacuum
- Approach#1: Traditional wire cut and free-falling ( Fig. 1 & 2 )
- Approach#2: Electromagnetic release and free-falling. (Fig 3 & 4 )
- The furnace can be modified into an HPCVD ( hybrid physical-chemical vapor deposition ) furnace via DIY a sample susceptor ( click picture below-left to see how )
- The furnace can be used for brazing of two flat < 1" diameter sample via top dead weight under vacuum at <1000° ( click picture below -right to see how )
|
|
Warning
Click here to learn more about the installation of a gas regulator.
|
- The tube furnaces with quartz tube are designed for use under vacuum and low pressure <0.2 bar / 3 psi / 0.02 MPa. Vacuum pressures may only be safely used up to 1000°C. The flow rate for gases should be limited to < 200 SCCM (or 200ml/min) for reducing thermal shocks to the tube.
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
|