Furnace Structure
|
-
Double-layer steel case with air cooling to ensure furnace surface temperature < 60 °C
-
Energy-saving high purity fibrous alumina insulation.
-
Heating Element: high-quality Fe-Cr-Al Alloy doped by Mo
- 86 " length x 5.5"dia heating chamber with 8 heating zone and 8 built-in thermal couples to accept processing tubes from 1" - 4" dia.x2700 mm length
- The splittable cover is lifted and closed by a pair of heavy-duty compressed air springs to make tube replacement easy.
|
|
Power & Voltage
|
- AC 208-240V Single Phase, 50/60 Hz
- 9 KW Max. (50A breaker required)
- Powder cord is included without plug
- Note: If use 208V, three-phase power panel, need min. 30A breaker
|
|
Continuous Temperature
|
- 1100 °C under flowing gas
- 1000 °C under vacuum < 1 torr
- 1000 °C under vacuum < 1 torr
|
|
Max. Heating Rate
|
<= 10 °C /min
|
|
Quartz tube & Thermal Block Placement
|
- Quartz Tube: 100 OD x 92 ID x2700 L mm
- Two fibrous alumina thermal blocks are included, which must be placed on two ends of the inside tube ( click pic below right )
|
|
|
- One pair stainless steel vacuum flange is included with a quick assembling flange (an outside clamp for quick installation) without cooling.
- Custom fitting is as request from customer ( 1/4NPT only , without valve )
|
|
Heating Zone
|
-
Eight independent heating zones: Left and right heating zone length: 200 mm each and the rest heating zone length 150 mm each
-
Total zone length: 60" ( 1500 mm ) ( Including spacers between heating zone)
- Max. Constant temperature zone 60" ( 1500 mm ) if 8 zones heated at the same temperature ( within +/- 5°C)
|
|
Temperature Controllers
|
- One touch screen digital temperature controller built-in a separated unit, which can control seven heating zones independently
- Each channel is PID automatically controlled and saves 10 programs and each program has max. 30 segments
- Built-in overheating & broken thermocouple protection.
- +/- 1 ºC temperature accuracy.
- Eight K-type thermocouples (one for each zone).
- RS485 Communications Port built-in. and software included,
- The furnace may be operated by PC ( please order PC separately ).
-
|
|
Compliance
|
- CE Certified
- The furnace is ready to pass NRTLor CSA certification at extra cost.
|
|
Warning
Click here to learn the installation of a gas regulator
|
- The tube furnaces with quartz/mullite tubes are designed for use under vacuum and low pressure < 0.2 bar
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
- * Vacuum pressures may only be safely used up to 1000°C
- The flow rate for gases should be limited to<200 SCCM (or 200ml/min) for reducing thermal shocks to the tube
|