Note
|
OTF-1200X-VII-5 is updated based on OTF-1200X-VII from Jan. 20, 2014
|
|
Furnace Structure
|
-
Double-layer steel case with air cooling to ensure furnace surface temperature < 60 °C
-
Energy-saving high purity fibrous alumina insulation.
-
Heating Element: 1300ºC grade Fe-Cr-Al Alloy doped by Mo
- 41 "length x 5.5"dia heating chamber with 7 built-in thermal couples to accept processing tube from 1" - 5" dia.x2000 mm length
- The splittable cover is lifted and closed by a pair of heavy-duty compressed air springs to make tube replacement at easy.
|
|
Power & Voltage
|
-
AC 208-240V Single Phase, 50/60 Hz
-
8.5 KW Max. (50A breaker required)
-
Powder cord is included without plug
|
|
Max. Temperature
|
1200 °C for < 1 hour under flowing gas
|
|
Continuous Temperature
|
1100 °C under flowing gas 1000 °C under vacuum < 1 torr
|
|
Max. Heating Rate
|
<= 10 °C /min
|
|
Quartz tube & Thermal Block Placement
|
-
Quartz Tube: 130 OD x122 ID x2000 L mm (5"x 79") (The tube is consumable , click picture below to order spare )
- Two fibrous alumina thermal blocks are included , which must be placed on two end of inside tube ( click pic below right )
|
|
Vacuum Flange
|
- One pair stainless steel vacuum flange is included with quick assembling flange ( a outside clamp for quick installation)
- Left flange installed with needle valve with 1/4" pipe port for gas inlet ( Pic below left )
-
Righ flange built in K25 port and Anti-Corrosive digital vacuum gauge to connect a vacuum pump
|
|
Heating Zone
|
|
|
Temperature Controllers
|
- One touch screen digital temperature controller built-in a separated unit, which can control seven heating zones independently
- Each channel is PID automatically controlled and saves 10 programs and each program has max. 30 segments
- Built-in overheating & broken thermocouple protection.
- +/- 1 ºC temperature accuracy.
- Seven K-type thermocouples (one for each zone).
- RS485 Communications Port built-in. and software included,
-
The furnace may be operated by PC ( please order PC separately ).
-
|
|
Optionals
|
- You may add a multi-channel gas delivery system under bottom table for CVD or atmosphere control applications ( click PIC below-left to order )
- You may consider a pressure control unit to keep constant pressure ( click picture below right to order )
|
Software & WiFi Control (Optional)
|
- Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30 days free trial) for immediate use.
-
Labview Based Temperature Control System (MTS01) enables users to edit temperature profiles, manage heat-treatment recipes, record and plot data for MTI furnaces.
- A wireless remote control provides up to 300 meters of operating range.
- The above features are available upon request at an extra cost (up to $1,000). Please contact us for more information.
|
|
Compliance
|
- CE Certified
- The furnace is ready to pass NRTLor CSA certification at extra cost.
|
|
Notice
|
|
|
Warning
Click here to learn the installation of a gas regulator
|
- The tube furnaces with quartz/mullite tubes are designed for use under vacuum and low pressure < 0.2 bar / 3 psi / 0.02 MPa
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
- * Vacuum pressures may only be safely used up to 1000°C
- The flow rate for gases should be limited to<200 SCCM (or 200ml/min) for reducing thermal shocks to the tube
|