Furnace Construction
|
- Double shell casing with three cooling fans
- High purity Alumina fibrous insulation for max. energy efficiency
- DC Gear motor drives the quartz tube with variable speed 0 - 10 RPM
- The furnace can be electrically tilted up to 30°
- The split cover enables faster cooling and easy operation.
- All electric components are UL or Met recognized and ready to pass any safety test worldwide
|
|
Heating Zone length
|
|
|
Processing tube & Flange
|
Four Mixing Blades Dimension
- One pair of 130mm SS sealing flanges with needle valves and a rotatable gas connector is included
- 60 OD straight quartz or SS310 steel tube is available upon request.
|
|
Gas Fittings
|
Inlet Fitting: Elbow Push-in connector for 6mm O.D. tube; Outlet Fitting: Elbow Push-in connector for 12mm O.D. tube |
|
Max. Working Temperature
|
-
1100°C for continuously
-
1200°C Max for < 1 hours
|
|
|
- Three programmable temperature controllers are included to control three heating zones independently
- temperature accuracy: +/- 1°C
- Three Omega K type 3mm OD x 6" L included
- The touch screen control panel is available upon request at an extra cost ($600). See Pic left
|
|
Tube Rotating Speed
|
0 - 10 RPM variable
|
|
Furnace tilt angle
|
0 - 30°
|
|
Chamber pressure
|
10-2 torr
|
|
Production yield
|
2 kg per batch
|
|
Vacuum and Filters
|
- One heavy-duty mechanical vacuum pump with an Exhaust Filter is included
- One cold vacuum trap is included to avoid CVD contaminating the vacuum pump.
-
|
|
Gas Delivery System
|
- Four-channel precision gas mixing and delivery system with Touch Screen is included for 1-4 kinds of gases CVD
- Note: If using flammable gas in the system, It must be placed in a fume hood with a suitable gas detector. You must get approved by professional safety personnel.
|
|
Power
|
4 KW Max
|
|
Voltage
|
208 - 240VAC, single-phase, (30 A Breaker required), 50/60Hz
|
| Warranty & Certificate |
One-year limited warranty (consumable parts such as processing tubes, o-rings, and heating elements are not covered by the warranty; replacements are available at related products below.) |
| Laptop, software & Control (Optional) |
- Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30-day free trial) for immediate use.
-
LabVIEW-based Temperature Control System (MTS01)enables users to edit temperature profiles, manage heat-treatment recipes, and record and plot data for MTI furnaces.
- A wireless remote control provides up to 300 meters of operating range.
-
|
| Compliance |
- CE Certified
- NRTL (UL61010) or CSA certification is available at extra cost.
|
|
Application Notes
|
- The continuous power feeding and processing can be realized by the use of Magnetic Fluid Sealed Swivel Coupling in line with the Volumetric Feeder and Rotary Furnace. Please view the assembling approach diagram and video below.
|
|
Warnings
|
- The tube furnace with quartz tube is designed for use under vacuum and low pressure < 0.2 bars / 3 psi
-
Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn about the installation of a gas regulator.
- Vacuum limit definition for all quartz tube furnaces: * Vacuum pressure may only be safely used up to 1000°C
|