Furnace Structure
|
- Double-layer steel structure with dual air cooling fans
- One pair of sliding rails are installed on the bottom of the furnace, which allows the furnace to be manually moved from side to side
- Vacuum flanges and a digital vacuum gauge are installed for immediate use.
- Sliding rail length is about 1200mm, the furnace can slide 600 mm distance
|
|
Power
|
2.5KW |
| Voltage |
AC 208-240V Single Phase, 50/60 Hz |
|
Max. Temperature
|
1200°C (<1 hour) |
|
Continuous Temperature
|
1100°C |
|
Heating& Cooling Rates
|
|
Heating Rate:
15°C/sec (RT - 150°C);
10°C/sec (150°C - 250°C);
7°C/sec (250°C - 350°C);
4°C/sec (350°C - 500°C);
3°C/sec (350°C - 550°C);
2°C/sec (550°C - 650°C); 1°C/sec (650°C - 800°C); 0.5°C/sec (800°C - 1000°C);
|
Cooling Rate:
15°C/sec (1000 - 950°C);
10°C/sec (950°C - 900°C);
7°C/sec (900°C - 850°C);
4°C/sec (850°C - 750°C);
2°C/sec (750°C - 600°C); 1.5°C/sec (600°C - 500°C); 1°C/sec (500°C - 400°C); 0.5°C/sec (400°C - 300°C);
|
Note: The max heating and cooling rate is achieved by pre-heat one side of the processing tube to the target temperature and then slide the furnace to the other side. You may choose to place the sample in the pre-heat area or not.
|
|
Heating Zone Length
|
440mm ( single zone )
|
|
Constant Temperature Zone
|
150mm (+/-3°C @ 1000°C)
|
|
Heating Elements
|
Fe-Cr-Al Alloy doped by Mo
|
|
Processing Tube
|
High purity fused quartz tube
O.D. : 3.2" (80 mm) I.D. : 3.0" (75 mm) Length: 55" (1400 mm)
Two foam ceramic blocks are included which must be placed inside tube to avoid heat radiation damaging O-ring
|
|
Temperature Controller
|
- PID automatic control with 30 programmable segments with +/ - 1°C accuracy
- Built-in over-temperature alarm and protection allow for operation without an attendant.
- Two K-type thermal couples are included. One is built in the furnace to control furnace temperature, and another inserted into the tube to monitor sample real temperature profile.
- One NIST certified thermometer is included to connect to the thermal couple inside the tube. the meter has RS485 Interface with Windows Software which can record temperature profile by PC.
- PC control module is included.
- Advanced Temperature Control (Optional)
- MTS02-Y Temperature control software kit (for YD518P series controllers) + 15" laptop package is available at the options bar. Click for detailed information about the MTS02-Y software kit.
- You may upgrade the temperature controller to the Eurotherm 3000 Temperature Controller w/ +/- 0.1ºC accuracy
|
|
Vacuum Flanges
|
- Stainless steel vacuum flanges are installed on both ends of the tube with heavy duty support.
- Right flange: flange, fittings for the vacuum gauge are included. The bellow, KF25 adapter, KF25 clamp are not included.
- Left flange: one needle valve and 1/4" thermocouple feed-through and digital vacuum gauge.
- An 24" length x 1/4" Dia. K-type thermocouple (included) connected to an Omega temperature calibrator (included) will be inserted through this flange to monitor the heating and cooling rates of the sample's position during furnace sliding.
- Click the pictures below to see details:
|
|
|
|
|
|
|
Right Flange
|
Left Flange
|
Temp. Calibrator
|
24" K-type T/C
|
Digital Vacuum Gauge
|
|
|
Vacuum Level
|
- Limited by the pump ( pump is not included, click on pictures to order )
- 10E-4 torr can be reached by the molecular pump
- 10E-2 torr by a mechanical pump.
|
|
Sliding Rails / Table
|
|
|
Application Notes
Click here to learn the installation of a gas regulator
|
- Tube furnaces with quartz tubes are designed for use under vacuum and low pressure < 0.2 bars / 3 psi / 0.02 MPa
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
- The flow rate for gasses should be limited to < 200 SCCM (or 200 ml/min) for reducing thermal shocks to the tube
- Vacuum limit definition for all quartz tube furnaces: * Vacuum pressures may only be safely used up to 1000°C
-
Click here to learn How to set up quartz/ ceramic tube and vacuum flange for MTI Tube Furnace.
- If you need a faster cooling jar please click the picture below to order a Portable Spot Air Cooling Conditioner.
- Warning: Fast sample cooling can be achieved by sliding the heating chamber away. Please wear protective gears such as heat resistant suit & gloves if you intend to do so. DO NOT TOUCH THE TUBE SURFACE WHEN IT IS HOT
|