Furnace Structure
|
|
Power & Voltage |
- AC 208-240V Single Phase, 50/60 Hz
- 30 KW Max. in tota
|
| Max. Temperature |
1100 °C (<1 hour) |
| Continuous Temperature |
1050 °C |
| Max. Heating Rate |
<= 20 °C /min |
Thermal Block Placement |
|
Heating Zone |
-
12 independent heating zones: Each heating zone length is 500 mm
- Total zone length: 6000 mm, or 20' in total
- Max. Constant temperature zone ~18‘ ’( 5500 mm ) if 6 zones are heated to the same temperature (within +/- 2 °C)
|
Temperature Controllers
|
- One touch screen digital temperature controller built-in a separated unit, which can control 12 heating zones independently
- Each channel is PID automatically controlled and saves 10 programs and each program has max. 30 segments
- Built-in overheating & broken thermocouple protection.
- +/- 1 ºC temperature accuracy.
- 12 K type thermocouples (one for each zone).
- RS485 Communications Port built-in. and software included,
- The furnace may be operated by PC ( please order PC separately ).
|
Tube & Flange |
- One 5" OD x 4.7" ID x 22‘ L 6500 mm) quartz tube is included
- One pair of stainless steel vacuum flange is included with a quick assembling flange (an outside clamp for quick installation). Left flange with needle valve with 1/4" pipe port for gas inlet and one pressure gauge; Right flange with KF 40 port for vacuum pump connection as shown below
-
|
Laptop, software & WiFi Control (Optional)
|
- Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30 days free trial)
-
Labview Based Temperature Control System (MTS01)enables users to edit temperature profiles, manage heat-treatment recipes, record and plot data for MTI furnaces.
- A wireless remote control provides up to 300 meters of operating range.
|
Compliance
|
- CE Certified
- NRTL or CSA certification is available upon request at the extra cost
|
Optional (click the picture to order at extra cost) |
- Please click here to order a digital vacuum gauge for precision vacuum control up to 10-4 torr
- If need high vacuum > 10-5 torr, please order molecular pump
- You may consider ordering a recirculating water chiller for water cooling
|
Application Notes
|
|
Warning
|
- Tube furnaces with quartz/mullite tube are designed for use under vacuum and low pressure <0.2 bar
- Attention: A two-stage pressure regulator must be installed on gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
- Vacuum pressures may only be safely used up to 1500°C
|