Specifications
| Furnace |
Structure |
Double-layer steel casing with air cooling. High purity alumina fiberboard liner for energy cost-effective operation. Three heating zones with three independent temperature controllers. Vacuum and gas mixing system are located in the bottom Mobile cart with 4 wheels. |
|||||||||
| Heating Element |
Fe-Cr-Al Alloy doped by Mo |
||||||||||
| Max. Heating Temp. |
1200°C |
||||||||||
| Continuous Heating Temp. |
1100°C |
||||||||||
| Max. Heating Rate |
<=20°C/min |
||||||||||
| Heating Zone Length |
Three zones: (880mm, 35" in total) Zone 1: 220 mm + Zone 2: 440 mm + Zone 3: 220 mm |
||||||||||
| Constant Temp. Zone |
20" (500 mm) length (three zones at equal temperature within +/- 1°C) |
||||||||||
| Tube Size and Materials |
Fused Quartz Tube 100 mm O.D x 92 mm ID. x 1400 mm |
||||||||||
| Temp. Controller |
Three programmable precision digital temperature controllers with 30 segments, which control each heating zone separately. PID adjustable with auto tune function to ensure temperature stability within +/ - 1 oC. Full protection in each controller for over-temperature and broken thermal couple. |
||||||||||
| Thermocouple |
Three K-type thermocouples |
||||||||||
| Power Input |
AC 208~240V, Single Phase, 50/60Hz, 7KW max. |
||||||||||
| Vacuum Fittings |
The Standard vacuum accessories include:
|
||||||||||
| HV Station High Vacuum Station |
Structure | Mobile cart size: 600 (L) x 600 (W) x 700(H), mm Max. Loading: 600 Lbs on top Molecular pump control Panel: LCD digital Inside: Pfeiffer vacuum pump Made in Germany Vacuum sensor is included |
|||||||||
| Vacuum sensor | The vacuum sensor has a KF-25 connector. High vacuum gauge (measuring range 10^-8 mbar ~ 10^3 mbar) |
||||||||||
| Control Panel |
One button operation - the diaphragm pump and the turbomolecular pump will start after pressing the "start" button. System controlled spinning speed - the controller will control the speed of the turbine spinning up and down according to the vacuum level and leaking rate of the reaction chamber. Self-protection - the system will protect the pump from damage by overheating and overcurrent if the leaking rate of the reaction chamber is too high to achieve the high vacuum. LCD screen - double lines parameters checking interface make the working easier to be handled.
|
||||||||||
| Volume Flow Rate | Nitrogen N2 33 L/s Helium He 39 L/s (2340L/minute) Hydrogen H2 32 L/s |
||||||||||
| Working Range | From 1000 mbar to <1E-7 mbar | ||||||||||
| Ultimate Pressure | 10^-5 torr (with tube furnace) |
||||||||||
| Power Input | 110V AC or 220V AC switchable, 50/60Hz | ||||||||||
| Power consumption | Duration/Max power: 100/110Watt | ||||||||||
| MFC Station Mass Flow Controller |
9 precision mass flow meters controllers are installed in the bottom mobile cart with the digital display panel of +/-1.5% of Full Scale.
|
||||||||||
| Optional Parts |
|
||||||||||
| Gas Pipe Connect |
|
||||||||||
| Shipping Info |
Dimension |
Furnace: 550 x 380 x 520mm MFC Station: 600 x 600 x 700mm HV Station: 600 x 600 x 700mm |
|||||||||
| Weight |
Net weight: 100Kg Shipping weight: 350 lb (two pallets) |
||||||||||
| Warranty |
One year limited warranty (Consumable parts such as processing tubes, o-rings, and heating elements are not covered by the warranty, please order the replacement at related products below.) |
||||||||||
| Compliance |
|
||||||||||
| Warning Click here to learn the installation of a gas regulator |
|
||||||||||
| Operation Instructions | Quick test instruction | ||||||||||