Power
|
- AC 208-240V Single Phase, 50/60 Hz
- 3 kW
|
|
Heating
|
- Maximum:1200< 0.5 hrs
- Continuous Working: 1100 or less
- Recommended heating rate ≤10/min
- Heating zone length: 400mm
|
| Thermocouple |
|
|
Specs
|
- Slide distance: 400mm (can be customized at extra cost)
- Adjustable speed range: 10-100mm/s (can be customized at extra cost) and can also be set to move back and forth
- The heating will automatically slide to the far right from the far left according to the temperature program after reaching a certain temperature
- The maximum heating rate for the left sample area: 3/s
- Rapid cooling rate for samples inside the furnace: 3/s (500-100), 0.7/s (500-200)
|
|
Temperature Controller
|
- Two PID automatic controllers with 50 programmable segments for control of heating & cooling rate, and dwell time.
- Built-in overheating & thermocouple broken protection.
- +/- 1 ºC temperature accuracy.
- DB9 PC Communications Port.
- PC Control:
- An MTS-02 control module is included for PC control on this furnace.
- A laptop PC can be ordered for $500 together with the furnace. We recommend ordering a computer (laptop) from us to avoid software installation problems as we will perform installation and testing before shipping. For more information please click here.
|
|
Vacuum Flanges
|
- One set of stainless steel sealed flanges
- The flanges and quartz tube are sealed with O-rings, pressure rings, and O-rings, and tightened by threaded compression to ensure sealing
- The left flange has a KF25 vacuum interface, needle valve, and barb venting port
- The right flange has a needle valve, G1/4 air inlet, and mechanical pressure gauge.
|
| Quartz Tube Options |
- 25 O.D x 19 I.D x 1400 L (mm)
- 50 O.D x 44 I.D x 1400 L (mm)
- 60 O.D x 55 I.D x 1400 L (mm)
- 80 O.D x 75 I.D x 1400 L (mm)
- 100 O.D x 95 I.D x 1400 L (mm)
|
|
Vacuum Level
|
|
|
Compliance
|
- CE Certified
- NRTL certification(UL61010) or CSA certification is available at extra cost.
|
|
Application Notes
|
- Tube furnaces with quartz tubes are designed for use under vacuum and low pressure < 0.12 MPa
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
- Vacuum pressures may only be safely used up to 1000°C
|