Features
|
- Mixing under vacuum for deaeration up to 1 Torr
- Two containers for 1000ml mixing or milling samples.
- Minimized contamination (no metal mixing blade)
- Thes speeds of main disk revolution and of container self-rotating are controlled independently upto 1200 RPM to provide various impact and shearing force.
- Sample Materials
- Nanopowders, ceramic powders
- Grease
- Slurry
- Paste
|
| Voltage Input |
- Single Phase AC 208 - 240V, 50/60 Hz
|
| Power |
|
| Container Capacity |
- Maximum volume for 1000mL container: 700mL
- Maximum total weight for 1000mL container: 1.2kg
|
Mixing and Speed
|
- Planetary Centrifugal Vacuum Mixing
- Main Platen Rotation (revolution motion)100 - 1200 RPM (Adjustable)
- Container Rotation (rotation motion)100-1200RPM (Adjustable whereas the container rotation speed can not be set lower than main platen rotation speed.)
|
| Vacuum Capability |
- The chamber is vacuum sealed with built-in mechanical pump.
- Pumping Rate: 165 L/min
- Ultimate Vacuum: -95 kPa
|
| Containers and Counter-Weights |
- Two standard containers and counter-weights are included
- Two 1000 mL containers with gas vent, made of high-density Polyethylene
- For the small amount of sample mixing, the adaptor for the samll container is available at extra cost.
- For ball milling, stainless steel liner and adaptor are available upon request at extra cost.
|
| Program |
- Store up to 6 programs in memory, as shown in Fig.1
- 3 programmable segments, as shown in Fig.2. For example:
- The maximum running time for each segment is 200 sec. The maximum total running time is 600 sec
- The rotation direction could be either clockwise (mode A) or counter-clockwise (mode B)
|
| Safety and Protection |
- Safety warnings and interlocks are integrated. The unit automatically stops when
- The unit is overloaded with the sample
- The sample is loaded in an unbalanced way
- The top cover is opened
|
| Application Notes |
- Always use the mixer on a level & stable surface for best performance and maximum safety
- Do not use the mixer in a hazardous atmosphere, or with flammable, explosive, corrosive, poisonous material! Check the chemical resistance of the container with different materials before loading the sample!!!
- Do not overload the mixer! Stop the mixing operation immediately if abnormal sound/vibration is observed
- Use a digital balance to make sure the sample loading is balanced before mixing
- Double-check if the containers are mounted correctly and securely before mixing
- Do not open the cover lid until the rotation has completely stopped! Do not retrieve the containers and/or put a hand close to the mixing fixture while the rotation has not completely stopped!
- Only use the included containers for the mixing operation! Double-check to make sure there is no foreign object in the mixer before mixing
- Please click underline to learn Centrifugal Force calculation.
|