Features
|
- 6 Channel precise Stream Metering Pump with upto 64 sample stations. (sample container/tube is mounted on the platform fixture/rack for program positioning)
- The source liquid container features liquid level detection;
- Auto washing function for the liquid transfer pipeline help avoid cross-contamination;
- The type and quantity of the required ingredients and the amount of each liquid added can be edit as recipes and programmed in the software.
|
| Working Voltage |
- 220VAC, single phase, 50Hz
- Power: 3kW
|
| Work Station |
XYZ triaxial module platform station
- X-axis: single guide slider module, the working distance is 700mm, built-in Servo motor;
- Y-axis: double guide slider module, the working distance is 600mm, built-in Servo motor;
- Z-axis: rail slider module, the working distance is 300mm, built-in Servo motor.
|
| 6 channel Metering Pump with Pipetting Tips |
- Mounting fixture/rack can accommodate upto 64 source containers/tubes
- 6 channels liquid pipetting heads work individually in sequence, each channel can deliver 0.1mL-20mL for each move.
- Pipetting Tips is easily replaceable
- A metering pump with PTFE liner can resist most of the acidic and alkaline based solutions.
- 6 liquid source tanks are made of SS 304, max volume is 3L.
- Delivery Reproducibility (C.V.%): ± 0.3%
|
| Sample Container |
- Cylindrical tube
- Adaptable size 2mL - 500mL
- Max size: Φ100mm x 100mm H
- Min size: Φ10mm x 30mm H
|
| Distribution Speed |
|
| Temperature |
- 25±3 (environmental temperature)
|
| Humidity |
|
| Cleanroom classification |
|
| Gas Source |
- Inert Gas Tank or Air Compressor (0.5~0.8Mpa) is needed for the machine operation, please purchase separately.
- Attention: If a gas cylinder is used instead of an air compressor(pic2), a two-stage pressure regulator(pic1) must be installed on the gas cylinder to limit the pressure within the required working range for safe operation. Click here to learn the installation of a gas regulator.
pic1 pic2
|
Programming & Sofware
|
- PC based data analysis software can save the dosing data of each sample tube automatically. Data can be exported in Excel format.
- HMI programming. Each step allows the user to set the pipette movement of each source and calibrate the positioning of the sample tube racks.
- Recipe maker allows the user to easily edit, store, and upload a protocol.
|