Input Power
|
- Single-phase 110 VAC or 208-240 VAC (selectable at Product Options), 50 / 60 Hz
- 700 W power
|
Control Panel
|
-
4.5" touch screen panel controls all necessary parameters for automatic operation, including:
- Purging pressure and time, as well cycle time ( -90KPa - 100Pa)
- Sealing mode: side / top-sealing without vacuum or Vacuum sealing
- Sealing Temperature: 50 ~ 260ºC Max. ( +/- 2 ºC) adjustable
- Recommended temperature: 200 for top sealing die, and 160 for bottom sealing die
- Vacuum standing time: 0 - 900 S Adjustable
- Heated sealing time: 0 ~ 99 seconds adjustable (3 ~ 4 seconds recommended)
|
Back Panel
|
- Inert gas port, compressive gas port, and vacuum ports are installed on the back panel
- You need an air pump to operate ( optional, click Pic. 1 to order)
- You need an inert gas tank to purge the chamber ( not included)
- The vacuum pump is required for final sealing ( optional, click pic 2 to order)
- Plastic tubes are included for immediate use
-
Pic.1 Pic. 2
|
| Vacuum Level |
- Vacuum range: -95 ~ 0 kPa adjustable by digital vacuum meter
- The vacuum pump is not included. Please click the picture to the left to order at MTI.
- If you have your own pump with KF25 PORT, a Barbed fitting to KF-25 Adapter is included
- The highest vacuum level is 10-2 torr by mechanical pump and 10-4-torr by turbopump
-
Click the picture to order a vacuum pump
|
| Sealing Die |
- Max battery size:
- Length ≤ 160 mm
- Width ≤ 180 mm
- Thickness ≤ 12 mm
- Sealing width: 4 mm (The sealing width could be customizable)
- Sealing length: ≤ 190 mm
- Sealing evenness: ≤ 0.03 mm overlapping precision of top and bottom sealing dies
- Sealing die type: Sealing blade with rubber buffer (soft-sealing)
|
Sealing Pressure
|
-
Pressure range: 0 ~ 7 kg/cm² adjustable. The vacuum chamber lid and sealing die need to be driven by compressed high-purity inert gasses with a recommended 60 psi pressure.
- Consumption: ~ 0.1 L compressed air per sealing
- Please use high purity inert gasses (Ar or N2), when operating in the glove box. Using Air Compressor driven when operating outside the glove box.
- Please note that the air compressor is not included and is not recommended for use in the glove box.
- Attention: If a gas cylinder is used instead of an air compressor(pic2), a two-stage pressure regulator(pic1) must be installed on the gas cylinder to limit the pressure within the required working range for safe operation.
|
Glove Box Compatibility
|
- A KF40 Feedthrough for Glovebox is included in the package.
- Please use high purity inert gasses (Ar or N2), when operating in the glove box. Using Air Compressor driven when operating outside the glove box.
- DO NOT set the pressure gauge above 90 psi (0.6 Mpa)!
|
| Compliance |
- CE Certified
- NRTL Certification is available upon request at an extra cost
|
| Application Notes |
-
Please wear gloves to operate the machine.
- Beware of the automatic vacuum chamber lid. Only one operator is allowed for this equipment.
- Please clean the sealing dies after each use. Clean the sealing die after it has cooled down to avoid corrosion by the electrolyte.
- Check the level and smoothness of the sealing die surface every week. If there is serious damage to the sealing die, please contact MTI to order a replacement.
- Lubricate the cylinder every week
- Check the screws in sealing dies and stage every month. Make sure they are tightened.
|