- Double-layer stainless steel structure for maximum safety
- The bottom loading mechanism by electric hydraulic lift which can hold 200 Kg Max. sample
- Water cooling front door and sealing jacket for max. service life.
- Touch-screen digital control panel for all parameters.
- Precision temperature control with +/- 3ºC uniformity inside the chamber and +/-1°C accuracy.
|
|
Inside Chamber Size
|
- Heated chamber: 600 x 600 x 600 mm ( ~ 24" x 24" x 24" . 216 Liter capacity )
- Energy-saving 1800-grade fibrous alumina insulation.
- High-purity Al2O3 coating.
|
|
Power
|
- AC 460V three-phase; 50/60 Hz
- 50 KW
- 63Amp 100Aair-breaker
|
|
Heating Element
|
- 12 Pcs Mo heating elements for 1700ºC. Max. or
- 12 Pcs SiC elements for 1400ºC. Max. or
-
-
|
|
Working Temperature
|
- Continuous: 1600°C or 1300 °C
- Max. 1700°C (< 1 hr) MoSi2 or 1400oC ( 1hr) SiC
|
|
Heating Rate
|
|
|
Temperature Accuracy
|
|
|
Thermocouple
|
two B-type thermocouples (control thermocouple and secondary alarm thermocouple)
|
|
Temperature Control
|
- Eurotherm controller with PID automatic control via SCR (Silicon Controlled Rectifier)
- 16 programmable segments for precise control of heating/cooling rate and dwell time.
- Over-temperature protection and thermocouple failure alarm functions allow operation without attendant(s).
- +/- 0.5ºC temperature accuracy.
- RS485 Communications Port
|
Gas Flow Control and
Vacuum Valve |
- Two flow meters (#1: 0~8L/min; #2: 0~4L/min) are installed on the front panel of the furnace.
- One heavy-duty vacuum pump is included for fast removal of oxygen from the heating chamber
-
|
Water Cooling
|
- A 16L/min recirculating water chiller is installed at the bottom of the furnace for cooling the furnace sealing flange (CE certified only)
- MET Certification (UL 1450) for the water chiller is available upon request at extra cost. Please contact our sales representatives for a quote.
|
|
Compliance
|
- CE Certified for the furnace, vacuum pump, and water chiller.
- NRTL (UL61010) or CSA certification is available at extra cost.
|
|
Application Notes
|
- For longer furnace life furnaces, we recommend the heating rate not to exceed 5 °C /min.
- The cooling rate shall also not exceed 5 °C /min.
- Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence and the cracks may be repaired with alumina coating.
- The refractory block must be inserted before closing door.
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
|
|
Warnings
|
- ATTENTION: Any damages caused by the use of corrosive and acidic gasses are not under the coverage of MTI One Year Limited Warranty.
- In order to prevent oxidation/corrosion damages to the Mo heating elements, it is necessary to keep a constant inert gas (Ar preferred) flow throughout the entire process. This is the most effective way to carry out any by-products.
|