Structure
|
- High-purity fibrous alumina insulation for maximum energy saving.
- Heating elements on all three sides of the interior chamber for fast heating and enhanced temperature uniformity.
- Double-layer steel structure with air cooling fan to keep the furnace's exterior safe to touch.
- The safety interlock will automatically power off the furnace when the door is opened.
- Built-in vent port (top) to remove undesirable contaminants and moisture and extend the lifespan of the heating elements and furnace.
|
Power
|
- 208~240V AC, Single phase, 50/60 Hz
- 7.5 kW, 40 A.
- Note: A power cord is included without a power plug.
A proper power plug is available via the figure on the left.
|
| Max Operating Temperature |
|
| Temperature Stability |
|
| Temperature Uniformity |
- +/- 8 °C @ 1000°C in the center 260 x 260 x 260 mm area
|
| Max. Heating Rate |
|
Temperature Controller
|
-
A digital temperature controller (Version 858) is included.
- Proportional–integral–derivative control (PID control) and auto-tune function
- 30 segments programmed with ramping, cooling and dwelling steps
- Built-in over-temperature alarm and thermocouple failure alarm
- +/- 1 ºC temperature control accuracy
- Default DB9 PC communication port
|
Temperature controller (optional)
|
- Upgrade a Eurotherm Temperature Controller for better temperature PID control with accuracy +/- 0.1oC
|
|
Laptop, software & WiFi Control (Optional)
|
- Brand new laptop with Microsoft Windows 11 for immediate use.
-
LabVIEW-based Based Temperature Control System (MTS03)enables the user to edit temperature profiles, manage heat-treatment recipes, and record and plot data for MTI furnaces.
- The above features are available upon request at an extra cost. Please contact us for more information.
|
| Gas Fitting & Exhaust |
- One 1/4" compression tube fitting as a gas Inlet is installed on the back panel of the furnace.
- One gas exhaust port is built on the top of the furnace casing.
|
Compliance
|
- CE Certified
- NRTL (UL61010) or CSA certification is available upon request at extra cost.
|
Application Notes |
- For longer furnace service life, we recommend the heating rate not to exceed 10 °C /min.
- The cooling rate shall also not exceed 10 °C /min.
- This furnace lacks vacuum sealing, and it is not recommended to use toxic or explosive gases in this furnace without the necessary safety controls and supervision.
- Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence, and the cracks may be repaired with alumina coating.
- You may order an alumina crucible by clicking the picture below.
-
Click here to choose a crucible
|