Structure
|
- High purity alumina fiber insulation surrounding the chamber provides maximum energy efficiency.
- Heavy-duty double-layer structure with a cooling fan to keep furnace exteriors safe to touch.
- Double-layer steel case with forced-air cooling to ensure lower safer face temperature.
- One venting port is built on the top of the furnace
- One air inlet is located at the back of the furnace.
- A separate programmable digital temperature controller for using the furnace inside the glove box or a remote location
- Ar gas is compatible when used in a glove box
|
| Power |
2.5KW (not suitable for running at 110V) |
| Max. Current |
15A |
| Voltage |
- 208 -240V Single Phase 50/60 Hz
- Note: Power cable included but without a plug. Please install a plug on your own or click the link below to choose a plug according to your country or your lab requirement.
- The feedthrough of a power cord for the glovebox is available upon request at extra cost.
|
| Operating Temp Range |
- 1250 °C < 1 hour
- 100-1200 °C continuously
|
| Stability at 1000 °C |
+/- 0.5 °C |
| Uniformity at 1000 °C |
+/- 5 °C at W×D×H 100×80×100 mm |
| Max. Heating Rate |
30°C/ min. |
| Thermocouple |
K types with Ceramic Sheath |
Temperature Controller
|
-
FA-YD518P-AG temperature controller is included.
- Proportional–integral–derivative control (PID control) and auto-tune function
- 30 segments programmed with ramping, cooling and dwelling steps
- Built-in over-temperature alarm and thermocouple failure alarm
- +/- 1 ºC temperature control accuracy
- Default DB9 PC communication port
- Note: you may use the multi-channel controller to control multi-furnace as an optional
|
Laptop, software (Optional)
|
- Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30 days free trial).
-
Labview Based Temperature Control System (MTS01)enables the user to edit temperature profiles, manage heat-treatment recipes, and record and plot data for MTI furnaces.
- A wireless remote control provides up to 300 meters of operating range.
|
Temperature controller (optional)
|
- MTS02-Y Temperature control software kit (for YD518P series controllers) + 15" laptop package is available at the options bar. Click for detailed information about the MTS02-Y software kit.
|
| Compliance |
- CE Certified
- NRTL (UL61010) or CSA certification is available at extra cost.
|
Application Notes
|
- For longer furnace service life, we recommend the heating rate not to exceed 10 °C /min.
- The cooling rate should not exceed 10 °C /min either.
- Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence and the cracks may be repaired with the alumina coating
- The heating module is easy to replace. Click the pic left to order the spare Heating module for replacement
- If firing material release vapor or smoking, please put the furnace into a fume hood as the left picture.
- Toxic, corrosive, or explosive gases are not allowed to use with this furnace without the necessary safety controls or supervision.
|