Power ( AC110V - 55V)
|
- Furnace input (two power inputs are required)
- 55VAC
- 110VAC or 220VAC (selectable from the option bar)
- 500W max. 50/60 HZ
- A transformer (110V or 220V - 55V) for running a single furnace under either 110V or 220VAC is included
- Two furnaces may be connected in series to run at 110VAC
|
|
Working Temperature
|
-
Max. Temperature: 1000 °C (< 1 hr)
-
Continuous working temperature: 900°C
-
Max. Heating Rate: 10°C /min
|
Heating Zone
|
|
|
Temperature Controller
|
-
PID automatic control and auto-tune function
- 30 programmable segments for precise thermal processing
-
Built-in protection for the over-heated and broken thermocouple.
-
Temperature Accuracy: +/- 1 °C
- K type thermocouple
|
|
Heating Elements
|
- Fe-Cr-Al Alloy doped by Mo
|
|
Protective Cover
|
- A specially made protective shield is mounted on the furnace for scald protection at high temperatures (800°C+).
|
Processing Tube
|
-
Fused quartz tube: 20 mm OD x 17 mm ID x 192 mm Length is included (Pic.1)
- Two ceramic tube blocks are included for blocking heat radiation generated from the tube center. (Pic.2)
- For high vacuum, please choose quartz block as Pic. 3
- The small crucible may be used for tube furnace.
-
Pic.1 Pic.2 Pic. 3 Pic.4
|
Vacuum Flange
|
- One end is KF25 quick clamp flange is included without valve
- Another end is screw tighten flange with 1/4 gas inlet
|
|
Vacuum Pressure
|
- 20 m torr can be achieved by the mechanical pump
-
Please click the picture below to order a vacuum pump
|
|
Dimensions
|
|
|
Shipping Weight
|
|
|
Warranty
|
-
One-year limited warranty with lifetime support (Consumable parts such as processing tubes, and O-rings are not covered by the warranty, please order replacements at related products below.)
|
Compliance
|
- CE Certified
- NRTL(UL61010) or CSA certification is available at extra cost.
|
| Operation Instructions |
|
| Application Note |
- The furnace is separable with a controller and easy to replace
- The furnace can be used a vertical position as the picture below
- The furnace is small enough to put into the glovebox through an airlock chamber
-
|