Input Power
|
- Input: AC220V±10%, three-phases, 50/60 Hz
- Total Power required: 8 KW (~32A)
|
|
Vacuum Chamber
|
- Stainless steel (SS304) chamber: 400mm ID. x 400 mm D x 450 mm H, polished finish.
- 100 mm observation window
- Hinged-type door for easy sample loading
- Motorized shutters for each evaporation boat,
- One ISO-100K port for connection with a turbo pump and a gate valve.
|
Vacuum Pump (optional)
|
- 260 L/s turbo molecular pump
- Vacuum level: 5.0x10-6 Torr ( after 40 minutes of pumping )
- Limit vacuum: 5.0x10-7 Torr ( after baking at 100 -150°C)
- Leaking rate: 5.0x10-7 Pa.
- Vacuum Port: DN 100 ISO-K
A mechanical pump (YTP-550-LD) is required as a backing pump to initially achieve the vacuum level of <1e-2 torr
|
Evaporation Source & Sample Holder
|
Four evaporation sources are included
- One tungsten heating boat with water-cooling without temperature control for high temperature coating up to 2000 °C. Tungsten heating boat, 70L x 12W x 3H (mm).
- Three resistance coil-heated crucibles for low-temperature coating up to 500 °C with precision temperature control
- Sample holder can accept a maximum sample of 150*150mm. Rotation speed: 0-30 rpm
- Evaporation distance from source to sample holder is up to 60 mm
- The sample stage with a heating function up to 700 °C
|
In-situ thickness monitor (optional)
|
- Quartz crystal microbalance is optional for thickness measurements
- The accuracy of thickness control is 1.0 Angstrom.
|
|
Control Unit
|
- Vacuum, thickness, and power control are integrated into one control box
- The Control Unit is separated from the vacuum chamber for easy moving
|
|
Coater in Glovebox (Optional)
|
|
|
Facility Requirement
|
- Space: 250 SF with height > 6 feet
- Chilled Water: > 16L/minute flowing water required for electrode cooling or using a recirculating water chiller
|
|
Application Notes
|
- Please check the fabrication process in detail below:
- https://mtixtl.com/collections/perovskite-solar-cells
|
|
Certificate
|
|