| Structure |
Standard System Structure Alternative System Structure (Please RFQ) |
||||||||||
| System Working Range | From the standard atmosphere to <E-4 mbar (under Cold Condition only; with special tube block) (Noted: Please don't use a high vacuum if the temperature is over 1500 °C) |
||||||||||
| Motorized Valve |
|
Upstream control Solution |
|||||||||
| Furnace | Furnace structure | Double-layer steel casing with air cooling. Two cooling fans inside to keep case temperature < 55°C | |||||||||
| Tube Materials | 99.8% High purity Al2O3 Ceramic | ||||||||||
| Tube Size | Please select the tube size in the options bar. 82mm O.D Alumina tube ---------- ID: 73mm; OD: 82 mm in diameter x 1000 mm length 101mm O.D Alumina tube ---------ID:92.10mm x OD: 101.60 mm x 1000 mm length. |
||||||||||
| Refractory lining | 1750 grade fiber alumina cast | ||||||||||
| Heating element | 1700 grade MoSi2 | ||||||||||
| Working Temperature | 1700°C (2912°F) Max. <1hr 1600°C continuously |
||||||||||
| Normal Working Temperature | 800 -1700°C (2732°F) | ||||||||||
| Temperature Controller |
MET certified 30 programmable segments for precise control of heating rate, cooling rate, and dwell time.
Built-in PID Auto-Tune function with overheating & broken thermocouple protection. Over-temperature protection and alarm allow for operation without attendant(s). +/- 1 ºC temperature accuracy. RS485 Communications Port.
Please click here to learn why SCR is more accurate than SSC (Solid state relay)
|
||||||||||
| Max. Heating / Cooling Rate | 5°C / min above 1200°C and 10°C/min below 1200°C | ||||||||||
| Heating Zone | 457 mm (18") | ||||||||||
| Constant Temperature Zone | 150 mm (6") (+/- 1°C at 1600°C ) | ||||||||||
| Temperature Accuracy | +/- 1°C | ||||||||||
| Power Input | Single Phase, 220V AC, 50/60 Hz | ||||||||||
| Power Output | Max. 5KW | ||||||||||
| High Vacuum Station | Structure | Mobile cart size: 600 (L) x 600 (W) x 700(H), mm Max. Loading: 600 Lbs on top Molecular pump control Panel: LCD digital Inside: Pfeiffer vacuum pump made in Germany Vacuum sensor is included |
|||||||||
| Vacuum sensor | Two vacuum gauges are installed in this system. 1. Mechanical vacuum gauge 2. High vacuum gauge (measuring range 10^-8 mbar ~ 10^3 mbar) is installed in the whole high vacuum system for monitoring the high vacuum status when operating the furnace with HV station support. |
||||||||||
| Control Panel | One button operation - the diaphragm pump and the turbomolecular pump will start after pressing the "start" button. System controlled spinning speed - the controller will control the speed of the turbine spinning up and down according to the vacuum level and leaking rate of the reaction chamber. Self-protection - the system will protect the pump from damage by overheating and overcurrent if the leaking rate of the reaction chamber is too high to achieve a high vacuum. LCD screen - double lines parameters checking interface make the working easier to be handled.
|
||||||||||
| Volume Flow Rate | Nitrogen N2 33 L/s Helium He 39 L/s (2340L/minute) Hydrogen H2 32 L/s |
||||||||||
| Working Range | From 1000 mbar to <1E-5 mbar (under Cold Condition only; Noted: Please don't use a high vacuum if the temperature is over 1500 °C) | ||||||||||
| Ultimate Pressure | <1E-6 mbar (without any leaking) | ||||||||||
| Power Input | 110V AC or 220V AC switchable, 50/60Hz | ||||||||||
| Power consumption | Duration/Max power: 100/110Watt | ||||||||||
| Compliance |
|
||||||||||
| Note |
|
||||||||||
| Operation Instructions |
|
||||||||||