Furnace Structure
|
The tube furnace has three layers
- Heating element center spiral type SiC filament ( 25 OD x 20 ID x 400 L mm )
- SiC sheath: quartz tube ID 30 × OD 36 mm × 622 mm.
- Sample layer: 36 ID × 58 OD × 622 mm L with vacuum sealed flange.
- Gas-cooled layer: 58ID × 90 OD × 622mm L to keep outside wall < 30ºC
customized size tube is available upon request
|
| Power |
- 208 - 240 VAC, single phase
- 2000W
|
|
Working Temperature
|
|
|
Heating Rate
|
- Max. Heating Rate: 10°C /min
|
| Heating Zone |
|
|
Temperature Controller
|
|
|
Vacuum Flange (Sold Separately)
|
- Special flange for three layers tube is included
- 1" flange is included for center tube
-
1" Flange Set
|
Fitting Ports (Sold Separately)
|
|
|
Vacuum Pressure
|
- Limited by vacuum pump
- 50 mtorr can be achieved by mechanical pump
- 10^-5 torr can be reached by molecular pump
|
|
Digital Pressure Gauge (optional)
|
A gas-type independent (above 10mbar) digital vacuum gauge with a wide range between 3.8x10-5 to 1125 Torr. Aside from greater measurement precision, this gauge reduces risks of chamber explosions caused by incorrect reading of gas pressures due to gas-type dependency. (Please click picture left for details) |
| Compliance |
- CE Certified
- All electric parts are UL / MET / CSA certified
- The furnace is ready to pass UL or CSA certification (UL61010 or SPE1000) on customer side at customer cost.
|
|
Application Notes
Click here to learn the installation of a gas regulator
|
- Tube furnaces with quartz tubes are designed for using under vacuum and low pressure < 0.2 bars / 3 psi / 0.02 MPa
- Attention: A two stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
- Vacuum limit definition for all quartz tube furnaces: * Vacuum pressures may only be safely used up to 1000°C
- The flow rate for gases should be limited to<200 SCCM (or 200ml/min) for reducing thermal shocks to the tube
- Please click here to see our typical temperature profile calibration results at various temperatures.
- In order obtain accurate temperature inside tube furnace, please click here to learn how to calibrate the temperature on your own.
- How to set up a ceramic tube and vacuum flange for MTI tube furnaces.
|