| Furnace structure |
-
Double-wall steel casing with forced air cooling.
-
Two cooling fans inside to keep the temperature of outside case < 70 ºC
-
High purity alumina fibrous insulation for max energy saving
-
One sliding rail to support flange to load sample with ease
-
Three heating zone to achieve a longer constant temperature zone or a certain temperature gradient
|
| Input voltage |
220VAC Single phase |
| power |
8kW |
| Fused quartz tube size |
- 8.5" O.D x 8.2" I.D x 51" Length
- One quartz tube is included, and the spare tube is available as an option ( click picture below to order )
|
| Heating element |
NiCrAl Resistance Wire |
| Working Temperature |
-
Max. heating temperature: 1100°C
-
Continuous working temperature: 400 -1000°C
-
Max. heating/cooling rate: 10°C/min
|
| Heating zone |
-
Total heating zone length 900 mm (36")
-
Three zones: 300 mm + 300 mm + 300mm (12" per zone)
-
Constant temperature zone: 500 mm (if three zones are in the same temperature)
|
| Temperature controller |
- Three precision temperature controller with PID function and 30 segments programmable
- One extra temperature monitor is built in to shut down power when temperature out of control by accident
- Double K type thermal couple installed.
- Temperature accuracy: +/-1°C
- Software and PC interface module is available at extra cost
- Eurotherm 3000 temp. controllers with +/- 0.1 oC accuracy are available upon request
Optional: PC interface & software Eurotherm temp. controllers
|
| Vacuum Flange and gas feedthrough ports |
Right side flange Left side flange Sliding flange
|
| Compliance |
- CE Certified
- NRTL or CSA certification is available upon request
|
| Warning |
- Tube furnaces with quartz tubes are designed for use under vacuum and low pressure < 0.2 bar / 3 psi / 0.02 MPa
-
Attention: Due to the high-pressure gas output from the gas cylinder, it is necessary to install the gas pressure regulator with it to reduce the output pressure, avoid overpressurizing the furance chamber and ensure a safe operation. Click here to learn the installation of a gas regulator.
- Vacuum limit definition for all quartz tube furnaces: * Vacuum pressures may only be safely used up to 1000°C
- The flow rate for gases should be limited to<200 SCCM (or 200ml/min) for reducing thermal shocks to the tube.
|