Furnace structure
|
- Double-layer steel casing with air cooling. The surface temperature of the case is < 70°C.
- Refractory fibrous alumina insulation liner for maximum energy saving.
- Heating element: NiCrAl Resistance Wire
-
Two fibrous ceramic tube blocks are included for blocking heat radiation from the inside tube. (Ceramic blocks must be fully inserted into the furnace chamber before heating)
- Hinged types of vacuum-sealed flanges are included to achieve a vacuum level of 10-5 torr by a turbo-pump
-
Holding 8" wafer
|
Heating zone |
- Total heating zone length 1200 mm
- Four zones: 300 mm + 300 mm + 300mm + 300 mm (24" per zone)
- Constant temperature zone:
-
1000 mm (if three zones are in the same temperature within +/- 3°C) longitude direction
- 1100 mm (if three zones are in the same temperature within +/- 5°C) longitude direction
- +/- 3 oC along the cross-section
|
Working temperature
|
- Max heating temperature: 1150°C ( < 1 hour )
- Normal working temperature: 200~1100°C
- Max heating & cooling rate: 5°C/min
- Temperature uniformity: please click the pic left to enlarge
|
Temperature control
|
|
| Processing Tube |
- High-quality quartz tube with OD 600 mm x ID 590 mm x L 1500 mm
|
Vacuum sealing & Flange
|
-
A pair of water cold SS vacuum sealing flange (Hinged Type) and 4 pieces of silicon O-ring are included.
-
Left Side Flange: Vacuum gauge, stainless steel needle valves, and 1/4" barbed hose fitting.
-
Right Side Flange: KF40 vacuum port with KF-40 Vacuum Right-Angle Valve for easy connection to KF40 bellows and pump. ( If using a pump with a KF25 port, you will need an adapter KF40-KF25.
- One 1/4" OD thermal couple insert port is built in the flange for easy temperature calibration.
- Vacuum Level: 10-2 torr by heavy-duty mechanical pump and 10-4 torr by regular turbopump. 5x10-6 torr can be reached by 700 L/s high-speed turbopumps and a modified flange adaptor
- Arecirculatingwaterchiller 16L/minis included for cooling flanges
|
|
Options for CVD
|
-
The tube furnace should be calibrated regularly to maintain temperature accuracy and uniformity. We are providing a simple calibration kit for you to complete this job at an extra cost The complete calibration kit includes (click the underline to see details):
- One 48"L ( 1/4"OD) K type calibration thermocouple
- One Omega temperature calibrator
- 5 feet long thermocouple power extension wire
-
You may consider ordering a turbopump for hi-vacuum up to 10-5 torr ( click pic 1 to order )
- For CVD application, you may need a gas delivery system as Pic. 2
- For flammable gases, especially H2, you may integrate a gas leak detecter into the control panel, as Pic.3
Pic. 1Pic. 2 Pc. 3
|
Rated Voltage
|
- Three Phase, 220V AC, 50/60 Hz
- Power consumption: ~50 KW
|
Compliance
|
- CE Certified
- NRTL or CSA Certification is available upon request, additional charges will apply
|
NOTE
|
|
|
Warning
|
- Tube furnaces with quartz tubes are designed for use under vacuum and low pressure < 0.2 bar / 3 psi
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn about the installation of a gas regulator.
- Vacuum pressures may only be safely used up to 1000°C
- The flow rate for gases should be limited to<200 SCCM for reducing thermal shocks to the tube
|