Features
|
- Simple Operation: With an understanding of Bragg's law, an individual can learn the machine operation within minutes
- Precision: Crystal orientation can be determined accurately within +/- 15" (0.0046°)
- Safety: Low X-Ray emission, automatic shutter, and lead foil cover reduce exposure to radiation
- Cost Effective: Features comparable to more expensive X-Ray diffractometers
|
|
Power Supply / Power
|
- Input AC 208 - 240V, 50Hz/ 60Hz selectable, 250W (50Hz is available upon request; the power plug is NOT included,)
- For 110V, 500W Transformer is required for this diffractometer
- The high voltage power supply is a discrete unit along with the diffractometer
-
|
|
X-Ray Tube
|
- Cu target, air cooling, Cu-Kα wavelength
- X-Beam size: 1 mm width x 10 mm length
- X-Ray tube with a monochromatic wave (Kα) filter improves accuracy
- Scintillation detector reduces harmful X-Ray radiation
- The X-Ray tube is a consumable part with 2 years warranty. The additional X-ray tube is available (See Table Standard Package of Machine below for price)
Monochromatic Filter Scintillation Tube
|
|
Max. Tube Current
|
|
|
X-Ray Detector
|
|
|
Measure Range Graduated Platform
|
-
2 Theta: 10° ~ 140°, Theta : 10° ~ 70°
|
|
Accuracy
|
- Digital display with measurement accurate within +/- 15" (0.0046°)
|
|
Sample Holder
|
- One spring load wafer holder is included for holding a flat wafer or substrate smaller than 2" (See Fig.1 Fig.2)
- One ceramic vacuum suction plate is included for holding a flat wafer or substrate larger than 2" (See Fig.3 Fig.4)
- A 6mm OD tube is required for the compression fitting on the back of the vacuum suction plate (See Fig.3)
Fig.1 Substrate Sample Holder Fig.2 Mounting Substrate Fig.3 Vaccum Suction Plate Fig.4 Large Wafer Example with Pump
|
|
Optional Parts
|
- A Vacuum Pump is required for using a vacuum suction plate
- A V-type Crystal Boule Holder is required for boule orientation measurement
-
- Vacuum Pump Boule Holder with Sample
|
|
Radiation Protection
|
- Built-in automatic shutter prevents operation with the cover opened
- Lead foil reinforced plastic cover reduces radiation leaking
- Warning: The cover must be CLOSED during operation
- The radiation level outside of the operating unit is measured as 0.07 uSv/h, which is the same as the background level, indicating good radiation protection from the unit
-
- safe radiation level outside the operating unit
|