| Power |
- 208-240 VAC, 50 A, 50/60 Hz, single phase
- 9 KVA
|
| High-Pressure Vessel |
- SS 304 double-layered water-cooled chamber, Ø294×652 mm
- Digital high-pressure transmitter, up to 25 MPa
- High-pressure solenoid valve as pressure relief control
- Mechanical pressure relief valve, preset at 22 MPa.
- 1/4" Gas inlet and outlet with high-pressure needle valves
- KF25 pneumatic solenoid valve for vacuum generation.
|
| Heating Chamber |
- Embedded Molybdenum resistance heating coil
- Heating area Ø80×130 mm
- Recommended sample size Ø60×60 mm
- Alumina thermal insulation sheathed by Mo jacket
|
| Temperature Control |
- 50 Segment programmable digital temperature controller is included
- Interlocks for over-temperature and thermocouple failure
- Temperature control accuracy: ± 1 °C
- B-type thermocouple
- RS-485 communication port ready for remote control.
|
| Working Temperature |
- 1500°C ≤8h
- 1400°C continuous
- Heating rate
- ≤ 2°C/min @ T<200°C
- ≤ 5°C/min @ 200°C ≤ T ≤1500°C
|
| Working Pressure |
|
| Working Gas |
|
| Touchscreen Panel |
- All-in-one control panel for temperature and pressure
|
|
Chain Hoist
(Optional)
|
- Manual chain hoist to lift upper flanges, insulation jacket, and heating chamber
- An optional electric chain hoist is available upon request.
- Load: up to 125 kg
- Power 208-240 VAC, single phase, 50/60 Hz, 750 W
|
| Consumables |
- O-rings for chamber seal, Ø195×3.5 mm (one-time use for high-pressure seal)
- O-rings for top cover plate, Ø150×7 mm (multiple use)
- Mo heating coil assembly
- Type B thermocouple
|
|
Vacuum Pump
(Optional)
|
- Vacuum pump is required for air evacuation but not included.
- MTI recommends using a dual-stage rotary vane pump
|
|
Water Chiller
(Optional)
|
- Chilled water circulation is required.
- MTI recommend using KJ6200 recirculating water chiller for best performance.
|
|
Gas Booster
(Optional)
|
- Gas booster is recommended for process at low-temperature but high-pressure (>10 MPa or 1500 psig) conditions due to insufficient gas pressure from a cylinder.
|
|
Air Compressor
(Optional)
|
- Compressed air is required @ 40 - 60 psig to drive the pneumatic solenoid valve
- MTI recommends using an air compressor or similar.
|
Compliance
|
- CE, NRTL (UL61010), or CSA certification is available at extra cost.
|
| Field Service |
- On-site installation, commissioning, and training are required but not listed in the unit cost
|